FDA Adverse Event Malfunction Summary report: N

NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE

MDR report key: 3790556 · Received May 5, 2014

Report

Report Number
0001811755-2014-01608
Event Type
Malfunction
Date Received
May 5, 2014
Date of Event
April 8, 2014
Report Date
April 9, 2014
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
FYD
PMA / PMN Number
K012991
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE TECHNICIAN WAS ABLE TO CONFIRM THE REPORTED EVENT. THE ROOT CAUSE WAS DETERMINED TO BE DUE TO THE POWER CORD DAMAGE. THE CORD WAS COMING OUT OF THE PLUG BECAUSE THE PLUG WAS DAMAGED. THE DEVICE WAS REPAIRED AT THE CUSTOMER SITE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE POWER CORD OF THE NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE WAS BROKEN WITH EXPOSED WIRES AND PARTIALLY MELTED DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES ASSOCIATED WITH THE DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE POWER CORD OF THE NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE WAS BROKEN WITH EXPOSED WIRES AND PARTIALLY MELTED DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES ASSOCIATED WITH THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
268886 NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE APPARATUS, EXHAUST, SURGICAL FYD STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1