M2A-MAGNUM PF CUP 54ODX48ID
Report
- Report Number
- 0001825034-2014-03531
- Event Type
- Injury
- Date Received
- May 5, 2014
- Date of Event
- April 23, 2014
- Report Date
- July 8, 2014
- Manufacturer
- BIOMET ORTHOPEDICS
- Product Code
- KWA
- PMA / PMN Number
- PK042037
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- ATTORNEY
Narratives
CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER POSSIBLE ADVERSE EFFECTS, NUMBER 1 STATES, ¿MATERIAL SENSITIVITY REACTIONS.¿ NUMBER 6 STATES, ¿INADEQUATE RANGE OF MOTION DUE TO IMPROPER SELECTION OR POSITIONING OF COMPONENTS.¿ NUMBER 14 STATES, ¿INTRAOPERATIVE OR POSTOPERATIVE BONE FRACTURE AND/OR POSTOPERATIVE PAIN.¿ NUMBER 15 STATES, ¿ELEVATED METAL ION LEVELS HAVE BEEN REPORTED WITH METAL-ON-METAL ARTICULATING SURFACES.¿ THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF¿S COMPLAINT AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED. THIS REPORT IS NUMBER 2 OF 4 MDRS FILED FOR THE SAME PATIENT (REFERENCE 1825034-2014-03530/03533).
THIS FOLLOW-UP REPORT IS BEING FILED TO RELAY ADDITIONAL INFORMATION, WHICH WAS UNKNOWN AT THE TIME OF THE INITIAL MEDWATCH.
LEGAL COUNSEL FOR PATIENT REPORTED THAT PATIENT UNDERWENT RIGHT TOTAL HIP ARTHROPLASTY ON (B)(6) 2010. PATIENT'S LEGAL COUNSEL FURTHER REPORTED PATIENT ALLEGATIONS OF PAIN, DISCOMFORT, SORENESS, DYSFUNCTION, LOSS OF RANGE OF MOTION, FLUID, DAMAGE TO BONE AND TISSUE, METAL POISONING, METALLOSIS AND ELEVATED METAL ION LEVELS. THERE HAS BEEN NO REPORTED REVISION PROCEDURE TO DATE. A REVIEW OF THE INVOICE HISTORY CONFIRMED THE INITIAL SURGERY DATE. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF¿S COMPLAINT AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED.
LEGAL COUNSEL FOR PATIENT REPORTED THAT PATIENT UNDERWENT RIGHT TOTAL HIP ARTHROPLASTY ON (B)(6) 2010. PATIENT'S LEGAL COUNSEL FURTHER REPORTED PATIENT ALLEGATIONS OF PAIN, DISCOMFORT, SORENESS, DYSFUNCTION, LOSS OF RANGE OF MOTION, FLUID, DAMAGE TO BONE AND TISSUE, METAL POISONING, METALLOSIS AND ELEVATED METAL ION LEVELS. A REVIEW OF THE INVOICE HISTORY CONFIRMED THE INITIAL SURGERY DATE. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF'S COMPLAINT AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED. ADDITIONAL INFORMATION RECEIVED FROM PATIENT OPERATIVE (OP) REPORT NOTES PATIENT WAS REVISED (B)(6) 2014 DUE TO PAIN. REVISION OP REPORT NOTES THE PRESENCE OF FLUID, DISCOLORATION OF SOFT TISSUES, AND CORROSION ON THE TRUNNION. THE CUP, MODULAR HEAD, AND TAPER INSERT WERE REMOVED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 267949 | M2A-MAGNUM PF CUP 54ODX48ID | PROSTHESIS, HIP | KWA | BIOMET ORTHOPEDICS | N/A | 023670 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 81 YR | Hospitalization| R |