FDA Adverse Event Malfunction Summary report: N

PROXIMATE LINEAR STAPLER

MDR report key: 3790266 · Received May 5, 2014

Report

Report Number
3005075853-2014-03027
Event Type
Malfunction
Date Received
May 5, 2014
Report Date
April 10, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS SHOWED THAT THE TL30 DEVICE ARRIVED IN GOOD VISUAL CONDITION AND WITH A RELOAD LOADED ON THE DEVICE. THE RELOAD WAS RETURNED VOID OF STAPLES. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT CYCLED, FIRED, AND ALL THE STAPLES FORMED AS INTENDED. THE DEVICE FIRED WITHOUT ANY DIFFICULTIES AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A ¿PARENCHYMAL RESECTION¿ PROCEDURE AFTER RELOADING THE DEVICE DID NOT STAPLE. NO PATIENT CONSEQUENCES REPORTED. PROCEDURE WAS COMPLETED WITH SAME LIKE DEVICE. NO FURTHER INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
267736 PROXIMATE LINEAR STAPLER STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA L4E543

Patients

Seq Age Sex Outcome Treatment
1 TR30 LOT: J4AZ3Y