PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2024168-2014-02802
- Event Type
- Injury
- Date Received
- May 5, 2014
- Date of Event
- April 10, 2014
- Report Date
- April 10, 2014
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4):DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION.THE CUSTOMER REPORTED THE DEVICE WAS DISCARDED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.
(B)(4). LOT NUMBER - SUBSEQUENT TO THE INITIAL MEDWATCH REPORT THE REPORTED LOT NUMBER WAS FOUND TO BE INCORRECT. LOT NUMBER HAS BEEN CHANGED FROM 30917K1 TO UNK. IT IS INDICATED THAT THE DEVICE WAS DISCARDED AND IS NOT RETURNING FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE CANNOT BE COMPLETED. REVIEWS OF THE LOT HISTORY RECORD AND COMPLAINT HISTORY OF THE REPORTED LOT COULD NOT BE CONDUCTED BECAUSE THE LOT NUMBER WAS NOT PROVIDED. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. MILD CALCIFICATION WAS REPORTEDLY PRESENT AT THE RIGHT COMMON FEMORAL ARTERY ACCESS SITE. THE PERCLOSE PROGLIDE DEVICE INSTRUCTIONS FOR USE STATE UNDER THE SPECIAL PATIENT POPULATIONS SECTION THAT THE SAFETY AND EFFECTIVENESS OF THE PERCLOSE PROGLIDE SMC DEVICES HAVE NOT BEEN ESTABLISHED IN PATIENT POPULATIONS WITH FEMORAL ARTERY CALCIUM, WHICH IS FLUOROSCOPICALLY VISIBLE AT THE ACCESS SITE.
IT WAS REPORTED THAT AFTER A DIAGNOSTIC CATHETERIZATION, ARTERIOTOMY CLOSURE OF A MILDLY CALCIFIED RIGHT COMMON FEMORAL ARTERY WAS ATTEMPTED USING A PERCLOSE PROGLIDE DEVICE. REPORTEDLY, A NEEDLE-TO-CUFF MISS OCCURRED AS THE NEEDLES DID NOT CAPTURE SUTURE. A SECOND PROGLIDE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WERE NO REPORTED ADVERSE PATIENT SEQUELAE AND NO REPORTED SIGNIFICANT CLINICAL DELAY IN THE PROCEDURE. THE OPERATOR WAS NOT KNOWN; THEREFORE, IT IS NOT KNOWN IF THE OPERATOR WAS TRAINED IN THE USE OF THE PERCLOSE PROGLIDE DEVICE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 267734 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | AV-TEMECULA-CT | 30917K1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | SHEATH: 6-FRENCH |