ULTRACISION HARMONIC ACE
Report
- Report Number
- 3005075853-2014-03033
- Event Type
- Malfunction
- Date Received
- May 5, 2014
- Date of Event
- February 28, 2014
- Report Date
- March 7, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
(B)(4). PIN HOLE THE DEVICE WAS RETURNED IN GOOD PHYSICAL CONDITION. THE DEVICE WAS FUNCTIONALLY TESTED ON THE GENERATOR AND THE MIN HAND ACTIVATION BUTTON WAS NOT FUNCTIONAL; WHEN TESTED WITH THE FOOTSWITCH AN ERROR CODE 5 SCREEN WAS DISPLAYED. THE DEVICE WAS DISASSEMBLED AND IT WAS FOUND THAT THE BLADE WAS CRACKED AT THE PIN HOLE UNDER THE SHROUD OF THE DEVICE; IN ADDITION, CORROSION WAS FOUND AT THE HAND ACTIVATION DOMES. IF THE HARMONIC SYSTEM SENSES AN INSTRUMENT FAULT DURING USE, AN AUDIBLE ALARM (TONE WITH LONG PULSES) WILL SOUND AND A VISUAL ALARM INDICATOR WILL APPEAR ON THE CONTROL PANEL. THE GENERATOR SYSTEM WILL REVERT TO STANDBY MODE, WHEN THE AUDIBLE ALARM (SOLID TONE) AND VISUAL ALARM INDICATOR APPEAR. ONCE MINOR BLADE DAMAGE HAS OCCURRED, SUBSEQUENT ACTIVATIONS MAY INCREASE DAMAGE SEVERITY AND RESULT IN AN ERROR CODE. THE CORROSION IN THE DOMES IS POSSIBLY CAUSED WHEN THE DEVICE WAS SOAKED FOR RE-USE OR THE DEVICE BEING SOAKED FOR CLEANING BEFORE SHIPMENT FOR ANALYSIS. THE ROOT CAUSE CANNOT BE IDENTIFIED BECAUSE IT IS UNKNOWN WHEN THE CORROSION OCCURRED. THE CORROSION WOULD HAVE INHIBITED ELECTRICAL CONTACT BETWEEN THE DOME AND THE CIRCUIT. OUR MANUFACTURING, STERILIZATION, PACKAGING AND SHIPMENT PROCESSES DO NOT INTRODUCE CORROSION TO THE DEVICE. IT IS PROBABLE THAT THIS MAY HAVE AFFECTED THE FUNCTIONALITY OF THE HAND ACTIVATION BUTTONS. WHEN THE BLADE IS DAMAGED AND PRIOR TO IT CRACKING, IT IS POSSIBLE TO FAIL PRE-RUN TEST, RESULTING IN AN ERROR CODE SCREEN. THIS WAS SEEN DURING OUR TESTING.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THERE WAS A SOLID TONE WITH ERROR CODE '5' AND THE SCALPEL COULD NOT PASS THE PRE-RUN TEST AGAIN. CHANGED TO ANOTHER DEVICE TO COMPLETE THE PROCEDURE. NO ADVERSE EVENT ON THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 266345 | ULTRACISION HARMONIC ACE | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | UNK | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR, HANDPIECE |