ENDOTAK RELIANCE
Report
- Report Number
- 2124215-2014-06698
- Event Type
- Injury
- Date Received
- May 5, 2014
- Date of Event
- March 18, 2014
- Report Date
- April 22, 2014
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- LWS
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). THE PRODUCT HAS BEEN RECEIVED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.
(B)(4). UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, ONLY THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED AND CAREFULLY EXAMINED. VISUAL ANALYSIS NOTED TRILUMEN INSULATION ABRASION 115-188 MILLIMETERS (MM) FROM THE TERMINAL PIN. THERE WERE SET SCREW MARKS ON THE TERMINAL PINS AND BLOOD OR BODY FLUID IN THE LUMENS. THE LEAD PASSED ELECTRICAL TESTING. DETAILED ANALYSIS REVEALED A SURFACE ABRASION WITH THE INSULATION ABRADED THROUGH TO THE DISTAL AND PROXIMAL HIGH VOLTAGE (HV) LUMENS 160-182 MM FROM THE TERMINAL PIN. WEBBING DAMAGE WAS NOTED BETWEEN ALL THREE LUMENS IN THIS AREA. THE DISTAL HV POLYTETRAFLUOROETHYLENE (PTFE) WAS WORN/ABRADED AS WELL AND THERE WAS MELTED METAL (ARCING DAMAGE) NOTED ON THE DBS CABLE. THE ABRASION WOULD MOST LIKELY HAVE BEEN LOCATED IN THE POCKET AREA. DUE TO THE LOCATION AND TYPE OF DAMAGE, IT APPEARS IT WAS MOST LIKELY CAUSED BY LEAD-ON-CAN CONTACT COUPLED WITH PRESSURE AND MOVEMENT WITHIN THE POCKET AREA.
THE LEAD IS EXPECTED TO BE RETURNED. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
ADDITIONAL INFORMATION OBTAINED INDICATES THAT THE SYSTEM WAS NOT TESTED WITH A PACING SYSTEM ANALYZER (PSA) AS THE LEAD WAS CLEARLY BROKEN. AS THE SURFACE OF THIS DEVICE WAS ALSO SLIGHTLY DISFIGURED, THIS LEAD TO THE PHYSICIAN TO DECIDE TO REPLACE BOTH COMPONENTS ON HAND. THE DEVICE WAS NOT TESTED AFTER REMOVAL AS IT WOULD HAVE REQUIRED THE USE OF A NEW LEAD FOR TESTING. THE PATIENT IS NOT PACEMAKER DEPENDENT.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) AND RIGHT VENTRICULAR (RV) LEAD EXHIBITED LOW OUT-OF-RANGE PACING LEAD IMPEDANCE MEASUREMENT LESS THAN 200 OHMS. SUBSEQUENTLY, UPON INSPECTION OF THE LEAD DURING EXTRACTION, IT WAS DISCOVERED THAT THE LEAD CONDUCTOR HAD FRACTURED AND THE INSULATION OF THE PROXIMAL END WAS DAMAGED. THE SYSTEM WAS EXPLANTED AND WAS SUCCESSFULLY REPLACED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
--
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 267671 | ENDOTAK RELIANCE | IMPLANTABLE LEAD | LWS | CPI - DEL CARIBE | 0185 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | Hospitalization| L| R | T175| F141| 0185 |