FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3790160 · Received May 5, 2014

Report

Report Number
2032227-2014-02244
Event Type
Malfunction
Date Received
May 5, 2014
Date of Event
April 14, 2014
Report Date
April 14, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ID
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED DURING THE BASIC OCCLUSION TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK. UNIT HAD BROKEN BELT CLIP SLOT, BATTERY TUBE THREADS, RESERVOIR TUBE LIP AND SCRATCHED DISPLAY WINDOW.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT SHE HAD HIGH BLOOD GLUCOSE OF 489 MG/DL AND THE INSULIN PUMP WAS NOT WORKING CORRECTLY. CUSTOMER STATED THAT HER INSULIN PUMP WAS SQUIRTING THE INSULIN OUT AND ALARMING DURING THE PRIMING PROCESS. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
267532 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-722NAP

Patients

Seq Age Sex Outcome Treatment