FDA Adverse Event Injury Summary report: N

DEXTRUS 4137

MDR report key: 3790104 · Received May 5, 2014

Report

Report Number
1028232-2014-01451
Event Type
Injury
Date Received
May 5, 2014
Date of Event
March 21, 2014
Report Date
April 23, 2014
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
NVN
PMA / PMN Number
P950037
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS THEREFORE BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ASSOCIATED WITH THE MANUFACTURE OF THIS PARTICULAR DEVICE. THE MANUFACTURING PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MANUFACTURING PROCESS WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MANUFACTURING.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR LEAD EXHIBITED AN IMPEDANCE INCREASE, INCREASED THRESHOLDS AND OCCASIONAL NOISE WITH OVERSENSING IN BIPOLAR. REPROGRAMMED TO UNIPOLAR AND NOISE ISSUE RESOLVED. THERE WERE NO ADVERSE PATIENT EFFECTS. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
267397 DEXTRUS 4137 PACER LEAD NVN BIOTRONIK SE & CO. KG 358764

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization