FDA Adverse Event Malfunction Summary report: N

7.5CM LARGE CRANIOTOME ATTCH

MDR report key: 3790089 · Received May 5, 2014

Report

Report Number
1045834-2014-11925
Event Type
Malfunction
Date Received
May 5, 2014
Date of Event
April 18, 2014
Report Date
April 18, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
PK011444
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION.  RELIABILITY ENGINEERING EVALUATED THE DEVICE.  AN ASSESSMENT WAS PERFORMED ON THE DEVICE WHICH FOUND THAT THE TIP WAS BROKEN OFF FROM IMPROPER ASSEMBLY OF THE ATTACHMENT ONTO THE DEVICE.  THEREFORE, THE REPORTED CONDITION WAS CONFIRMED.  THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE MISUSE, ABUSE, AND OR ERROR.  IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING STERILE PROCESSING, IT WAS DISCOVERED THAT THE TIP OF THE CRANIOTOME DEVICE WAS ¿BROKEN OFF¿. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
267392 7.5CM LARGE CRANIOTOME ATTCH MOTOR, DRILL, ELECTRIC - CRANIOTOME HBC DEPUY SYNTHES POWER TOOLS NA

Patients

Seq Age Sex Outcome Treatment
1