FDA Adverse Event Malfunction Summary report: N

MEGA NEEDLE DRIVER INSTRUMENT

MDR report key: 3790007 · Received May 5, 2014

Report

Report Number
2955842-2014-02756
Event Type
Malfunction
Date Received
May 5, 2014
Date of Event
April 8, 2014
Report Date
April 9, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS FOUND THE INSTRUMENT'S GRIP CABLE TO BE BROKEN AT THE DISTAL IDLERS. THE IDLER PULLEY SPUN FREELY AND DID NOT EXHIBIT ANY DAMAGE. THE CABLE SEGMENT WAS OBSERVED TO BE STICKING OUT AT THE WRIST OF THE INSTRUMENT. THE CABLE BROKE UNDER TENSILE LOADING. THE OTHER CABLES AT THE WRIST OF THE INSTRUMENT WERE UNDAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT IN ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED BROKEN CABLE IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI HYSTERECTOMY SURGICAL PROCEDURE, BROKEN CABLES WERE IDENTIFIED AT THE DISTAL END OF THE MEGA NEEDLE DRIVER INSTRUMENT. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO A PATIENT. THERE WAS NO PATIENT HARM, ADVERSE OUTCOME OR INJURY REPORTED.