DUAL SPEED CEMENT INJECTION GUN
Report
- Report Number
- 0001811755-2014-01602
- Event Type
- Malfunction
- Date Received
- May 5, 2014
- Date of Event
- April 11, 2014
- Report Date
- April 11, 2014
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- JDZ
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
A FOLLOW UP REPORT WILL BE FILED AFTER THE QUALITY INVESTIGATION HAS BEEN COMPLETED.
BASED ON THE MANUFACTURING ENGINEER¿S EVALUATION OF THE DEVICE, THE CAUSE WAS DETERMINED TO BE THE ROD CATCHING ON THE BUSHINGS DUE TO INTERFERENCE. THE DEVICE WAS REPAIRED AND RETURNED TO THE USER FACILITY.
DURING INSPECTION AT THE USER FACILITY THE DUAL SPEED CEMENT INJECTION GUN WAS PRODUCING METAL SHAVINGS. IT WAS REPORTED THAT THERE WAS NO ASSOCIATED PROCEDURE. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.
DURING INSPECTION AT THE USER FACILITY, THE DUAL SPEED CEMENT INJECTION GUN WAS PRODUCING METAL SHAVINGS. IT WAS REPORTED THAT THERE WAS NO ASSOCIATED PROCEDURE. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.