FDA Adverse Event Malfunction Summary report: N

ANGLED ACET INSERTR

MDR report key: 3789271 · Received May 5, 2014

Report

Report Number
1818910-2014-18017
Event Type
Malfunction
Date Received
May 5, 2014
Date of Event
April 29, 2014
Report Date
April 29, 2014
Manufacturer
DEPUY ORTHOPAEDICS, INC. 1818910
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION CONFIRMED THAT THE LOCKING CATCH OF THE ANGLED ACET INSERTER HAD FRACTURED. PREVIOUSLY A CORRECTIVE ACTION WAS IDENTIFIED FOR THIS FAILURE MODE WHEREBY A DESIGN CHANGE WAS IMPLEMENTED AND ROUTED ON ECO-193427 (DWG-106984) IN 2007 IN ORDER TO STRENGTHEN THE LOCKING CATCH. THE DESIGN VERIFICATION WAS COMPLETED AND ROUTED ON DVA-101743-FDE AND CONCLUDED THAT THERE ARE NO OUTSTANDING ACTIONS IDENTIFIED. THIS PRODUCT WAS MANUFACTURED PRIOR TO THE IMPLEMENTATION OF ECO-193427. THE COMPLAINT SHALL BE CLOSED WITH A JUSTIFIED CONCLUSION IT WILL BE ENTERED INTO THE COMPLAINT DATABASE AND MONITORED THROUGH TREND ANALYSIS.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Description of Event or Problem · 1

PLASTIC SLIDE LOCKING MECHANISM BROKE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
269102 ANGLED ACET INSERTR HIP INSTRUMENT/TRIAL LXH DEPUY ORTHOPAEDICS, INC. 1818910 P00806001

Patients

Seq Age Sex Outcome Treatment
1