FDA Adverse Event Malfunction Summary report: N

THERMOPHORE AUTOMATIC HEAT PACK

MDR report key: 3788602 · Received April 1, 2014

Report

Report Number
1811605-2014-00078
Event Type
Malfunction
Date Received
April 1, 2014
Date of Event
March 10, 2014
Report Date
April 1, 2014
Manufacturer
BATTLE CREEK EQUIPMENT
Product Code
IRT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

OUR INSPECTION FOUND A SMALL CUT IN THE CORD NEAR THE SWITCH THAT COULD HAVE EXPOSED THE USER TO HAVE WIRE. THE INSPECTION ALSO REVEALED CUSTOMER MISUSE AS OUR INSPECTION FOUND THAT: POLY DIRTY OR STAINED, CORD TWISTED, CORD CUT/BURNED STRAIN RELIEF, SWITCH OK, PAD BUNCHED, BENT/BROKEN LEAD, BENT/BROKEN THERMOSTAT, THERMOSTAT DISCOLORATION. THIS TELLS US THAT THE PAD WAS NOT BEING PROPERLY USED AS PER OUR INSTRUCTIONS IN THE MANUAL. CORD BREAKAGE USUALLY OCCURS WHEN THE CORD IS REPEATEDLY OVER BENT NEAR THE SWITCH. WE HAVE INITIATED A CAPA PROJECT (B)(4) TO REDUCE THE OCCURRENCE OF THIS ISSUE. A JEW JACKETED CORD DESIGN WAS LAUNCHED IN PRODUCTION ON 03/10/2014 THAT CLOSES (B)(4) PENDING A FUTURE EVAL OF THE EFFECTIVENESS OF THE SOLUTION.

Description of Event or Problem · 1

THE CUSTOMER STATED THE CORD CAUGHT FIRE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
196303 THERMOPHORE AUTOMATIC HEAT PACK ELECTRIC MOST HEAT PACK IRT BATTLE CREEK EQUIPMENT 055

Patients

Seq Age Sex Outcome Treatment
1