THERMOPHORE AUTOMATIC HEAT PACK
Report
- Report Number
- 1811605-2014-00078
- Event Type
- Malfunction
- Date Received
- April 1, 2014
- Date of Event
- March 10, 2014
- Report Date
- April 1, 2014
- Manufacturer
- BATTLE CREEK EQUIPMENT
- Product Code
- IRT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
OUR INSPECTION FOUND A SMALL CUT IN THE CORD NEAR THE SWITCH THAT COULD HAVE EXPOSED THE USER TO HAVE WIRE. THE INSPECTION ALSO REVEALED CUSTOMER MISUSE AS OUR INSPECTION FOUND THAT: POLY DIRTY OR STAINED, CORD TWISTED, CORD CUT/BURNED STRAIN RELIEF, SWITCH OK, PAD BUNCHED, BENT/BROKEN LEAD, BENT/BROKEN THERMOSTAT, THERMOSTAT DISCOLORATION. THIS TELLS US THAT THE PAD WAS NOT BEING PROPERLY USED AS PER OUR INSTRUCTIONS IN THE MANUAL. CORD BREAKAGE USUALLY OCCURS WHEN THE CORD IS REPEATEDLY OVER BENT NEAR THE SWITCH. WE HAVE INITIATED A CAPA PROJECT (B)(4) TO REDUCE THE OCCURRENCE OF THIS ISSUE. A JEW JACKETED CORD DESIGN WAS LAUNCHED IN PRODUCTION ON 03/10/2014 THAT CLOSES (B)(4) PENDING A FUTURE EVAL OF THE EFFECTIVENESS OF THE SOLUTION.
THE CUSTOMER STATED THE CORD CAUGHT FIRE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 196303 | THERMOPHORE AUTOMATIC HEAT PACK | ELECTRIC MOST HEAT PACK | IRT | BATTLE CREEK EQUIPMENT | 055 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |