FDA Adverse Event
Malfunction
Summary report: N
PRIME CARE TRANSCEND ADVANCED MATTRESS
MDR report key: 3788555
·
Received April 3, 2014
Report
- Report Number
- 3007538326-2014-00023
- Event Type
- Malfunction
- Date Received
- April 3, 2014
- Date of Event
- March 12, 2014
- Report Date
- April 3, 2014
- Manufacturer
- PRIMUS MEDICAL, LLC
- Product Code
- IKY
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
UPON INSPECTION OF THIS MATTRESS, THE URETHANE COVER BUBBLED AT THE FOOT END OF THE MATTRESS COVER BUT DID NOT PEEL AWAY EXPOSING THE FOAM OF THE MATTRESS. THIS MATTRESS HAS BEEN ISOLATED IN THE NON-CONFORMING ROOM UNTIL INVESTIGATION IS COMPLETE AND MATTRESS HAS BEEN DISPOSITIONED. A NEW MATTRESS WAS DELIVERED TO THE FACILITY ON (B)(4) 2014. THIS PROBLEM HAS BEEN ASSIGNED TO CAPA (B)(4), AND A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE CORRECTIVE ACTION.
Description of Event or Problem · 1
MATTRESS COVER IS DELAMINATING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 202186 | PRIME CARE TRANSCEND ADVANCED MATTRESS | IKY | PRIMUS MEDICAL, LLC | TACM3580 | 61312 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |