FDA Adverse Event Malfunction Summary report: N

I-STAT

MDR report key: 378603 · Received February 19, 2002

Report

Report Number
MW1024148
Event Type
Malfunction
Date Received
February 19, 2002
Date of Event
January 15, 2002
Report Date
February 19, 2002
Manufacturer
I-STAT
Product Code
CHL
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
WA, US
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

MARKED INCREASE IN FAILURE RATE OF I-STAT CARTRIDGES OBSERVED IN LAST MONTH, SUBSTANTIATED ON 1/29/2002. CO NOTIFIED THAT LOT A01159 6+ CARTRIDGES FOR THE POINT OF CARE I-STAT INSTRUMENT WERE GIVING NO DATA FOR POTASSIUM ON UP TO 17 % OF ATTEMPTS (BACKGROUND 1-2 % HERE). DEVICE USED IN A VARIETY OF PT SETTINGS. MOST OF DATA IS CAPTURED IN A DATA MANAGEMENT SYSTEM AND HAS BEEN DOWNLOADED TO THE CO. FAILURE RATE NOT TRACED TO ANY PARTICULAR PT POPULATION. HOSP IS A HIGH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 I-STAT POINT OF CARE BLOOD GAS AND ELECTROLYTE TEST SYSTEM CHL I-STAT 6 + CARTRIDGE A01189

Patients

Seq Age Sex Outcome Treatment
1 * Other