ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-12063
- Event Type
- Malfunction
- Date Received
- April 30, 2014
- Report Date
- April 20, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1: DATE OF SUBMISSION 05/23/2014. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 05/19/2014 WITH THE FOLLOWING FINDINGS: CS 078-0008 VERIFIED IN BLACK BOX AND HISTORY. BATTERY COMPARTMENT CRACKED FROM CASE SEAL TO GRIP. RECEIVED CS 078-0008 DURING INVESTIGATION. OPENED PUMP AND REMOVED FROM CASE. CORROSION ON MOTOR FLEX CABLE CONNECTOR.
THE REPORTER CONTACTED ANIMAS ON (B)(4) 2014 REPORTING A CALL SERVICE ALARM ISSUE. THERE IS NO REPORTED ADVERSE EVENT WITH THIS COMPLAINT. THIS REPORT IS MADE BASED ON THE ALLEGATION OF THE CALL SERVICE ALARM ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 260867 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 32 YR |