FDA Adverse Event Malfunction Summary report: N

G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 3782333 · Received April 30, 2014

Report

Report Number
3004753838-2014-06907
Event Type
Malfunction
Date Received
April 30, 2014
Date of Event
April 1, 2014
Report Date
April 8, 2014
Manufacturer
DEXCOM, INC.
Product Code
MDS
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION OF THE RETURNED RECEIVER AND REVIEW OF THE DATA LOG CONFIRMED THE REPORTED OUT OF RANGE SIGNAL. ADDITIONALLY, A TRANSMITTER (PART NUMBER: 9438-01; SERIAL NUMBER (B)(4)) WAS RETURNED AND EVALUATED. THERE WERE NO DEFECTS FOUND WITH THE TRANSMITTER.

Description of Event or Problem · 1

DISTRIBUTOR CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2014 TO REPORT AN OUT OF RANGE SIGNAL ON (B)(6) 2014. PATIENT DID NOT REPORT ANY INJURY OR MEDICAL INTERVENTION AT THE TIME OF CONTACT WITH DEXCOM TECHNICAL SUPPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
259941 G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM MDS MDS DEXCOM, INC. MT20649 5086029

Patients

Seq Age Sex Outcome Treatment
1 Other