FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 3782083 · Received April 30, 2014

Report

Report Number
1416980-2014-13990
Event Type
Malfunction
Date Received
April 30, 2014
Date of Event
April 8, 2014
Report Date
April 8, 2014
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE HOMECHOICE (HC) DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB (PAL). THE EVENT HISTORY LOG REVIEW SHOWED NO KEYSTROKES, PROGRAMMING, OR USE RELATED EVENTS THAT INDICATED AND/OR CONTRIBUTED TO THE REPORTED ISSUE. THE EXTERNAL/INTERNAL INSPECTION WAS PERFORMED AND THE DEVICE PASSED, ALONG WITH ALL OF THE ELECTRICAL TESTING. THE TEMPERATURE WAS VERIFIED TO BE WITHIN SPECIFICATIONS. DURING THE EVALUATION THE FLUID WAS TRANSFERRED ABOVE THE RITE (RETURNED INSTRUMENT TEST EVALUATION) SPECIFIED LIMITS, AND THE VOLUMETRIC ACCURACY TEST THAT FOLLOWED SHOWED THAT THE DEVICE HAD FAILED. PAL PERFORMED A MORE DETAILED INSPECTION OF THE DOOR ASSEMBLY. THE INSPECTION SHOWED THAT THE PISTON FOAM WAS DETERIORATED. THE RESULTS OF THE EVALUATION REVEALED THE CAUSE OF THE FAILURE TO BE THE DETERIORATED PISTON FOAM. THE PISTON FOAM WAS TO BE SCRAPPED AND THE DEVICE WAS SENT FOR SERVICING. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE (HC) DEVICE, A BAXTER TECHNICIAN DETERMINED THE HC MACHINE FAILED FLUID VOLUME ACCURACY TESTING. DEVICE FAILED DURING EVALUATION, NO PATIENT INVOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
260226 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1