FDA Adverse Event Malfunction Summary report: N

ESPRIT VENTILATOR

MDR report key: 3781979 · Received April 30, 2014

Report

Report Number
2031642-2014-00304
Event Type
Malfunction
Date Received
April 30, 2014
Report Date
April 22, 2014
Manufacturer
RESPIRONICS CALIFORNIA, INC.
Product Code
CBK
PMA / PMN Number
K981072
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Description of Event or Problem · 1

FACTORY ANALYSIS OF A RETURNED POWER SUPPLY REVEALED A MALFUNCTION THAT MAY RESULT IN THE UNIT SHUTTING DOWN DURING OPERATION WHEN AC POWER IS RESTORED. IF A FAILURE OF THE POWER SUPPLY OCCURS DURING USE AND THE VENTILATOR SHUTS DOWN DUE TO A LOSS OF AC POWER, AN AUDIBLE POWER FAIL ALARM WILL ACTIVATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
258888 ESPRIT VENTILATOR VENTILATOR, CONTINUOUS CBK RESPIRONICS CALIFORNIA, INC. V1000 NA

Patients

Seq Age Sex Outcome Treatment
1