FDA Adverse Event
Malfunction
Summary report: N
DURATA STS OPTIM ACTIVE FIXATION, DF4 CONNECTOR
MDR report key: 3781922
·
Received April 30, 2014
Report
- Report Number
- 2938836-2014-09908
- Event Type
- Malfunction
- Date Received
- April 30, 2014
- Date of Event
- February 14, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
NO COMPLAINT RECEIVED WITH RETURN OF DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. EXTERNAL INSULATION ABRASION WAS NOTED AT 23.8-24.4CM FROM THE CONNECTOR PIN, CONSISTENT WITH LEAD FRICTION TO ANOTHER DEVICE OR FEATURE OF THE HEART. THE SILICONE INSULATION WAS INTACT AT THIS LOCATION.
Description of Event or Problem · 1
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 259374 | DURATA STS OPTIM ACTIVE FIXATION, DF4 CONNECTOR | DEFIBRILLATION LEAD | NVY | ST. JUDE MEDICAL, INC., CRMD | 7121Q/58 | NA |