FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION, DF4 CONNECTOR

MDR report key: 3781918 · Received April 30, 2014

Report

Report Number
2938836-2014-09998
Event Type
Injury
Date Received
April 30, 2014
Date of Event
March 27, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A FOLLOW UP, LEAD DISLODGEMENT DUE TO TWIDDLERS SYNDROME WAS OBSERVED. LEAD TESTS SHOWED ABNORMAL SENSING, PACING AND IMPEDANCE MEASUREMENTS. THE LEAD WAS EXPLANTED AND REPLACED. PATIENT WILL BE MONITORED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
258809 DURATA STS OPTIM ACTIVE FIXATION, DF4 CONNECTOR DEFIBRILLATION LEAD NVY ST. JUDE MEDICAL, INC., CRMD 7122Q/58 NA

Patients

Seq Age Sex Outcome Treatment
1 69 YR Required Intervention