FDA Adverse Event Malfunction Summary report: N

ARCHITECT ESTRADIOL

MDR report key: 3781890 · Received April 30, 2014

Report

Report Number
3005094123-2014-00011
Event Type
Malfunction
Date Received
April 30, 2014
Date of Event
March 31, 2014
Report Date
April 7, 2014
Manufacturer
A.I.D.D LONGFORD
Product Code
JLW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Additional Manufacturer Narrative · 1

NO PATIENT SAMPLE OR PRODUCT RETURN WAS RECEIVED. TESTING WAS PERFORMED USING A RETAINED KIT OF ARCHITECT ESTRADIOL LOT 34908UI00. CONTROLS AND PANELS WERE EVALUATED. ALL VALIDITY AND ACCEPTANCE CRITERIA WERE MET. A REVIEW OF HISTORICAL TESTING WAS PERFORMED WHICH INDICATED THAT LOT 34908UI00 PASSED ALL THE REQUIRED SPECIFICATIONS. A REVIEW OF ALL NON-CONFORMANCE REPORTS AND DEVIATIONS ASSOCIATED WITH ARCHITECT ESTRADIOL REAGENT LOT 34908UI00 AND ALL ASSOCIATED SUB-LOT NUMBERS DEMONSTRATED THAT THERE HAVE BEEN NO KNOWN QUALITY ISSUES IN RELATION TO THIS PRODUCT SINCE MANUFACTURE. CUSTOMER COMPLAINT DATA WAS REVIEWED AND NO ADVERSE TRENDS WERE IDENTIFIED. THE ARCHITECT ESTRADIOL REAGENT PACKAGE INSERT WAS REVIEWED AND WAS FOUND TO ADEQUATELY ADDRESS THE ISSUE. THE INVESTIGATION DID NOT IDENTIFY A MALFUNCTION / DEFICIENCY.

Description of Event or Problem · 1

THE CUSTOMER STATED THAT DECREASED ARCHITECT ESTRADIOL RESULTS WERE GENERATED. AN INITIAL RESULT OF 2657 PMOL/L WAS GENERATED WITH SID (B)(6). A NEW SAMPLE (SID (B)(6)) 2 DAYS LATER GENERATED AN ESTRADIOL RESULT OF 807 PMOL/L. THE TWO SAMPLES WERE REPEATED AND GENERATED SIMILAR RESULTS. SID (B)(6): 2562 PMOL/L; SID (B)(6): 810 PMOL/L. THERE WAS NO REPORT OF IMPACT TO PATIENT MANAGEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
259392 ARCHITECT ESTRADIOL ESTRADIOL JLW A.I.D.D LONGFORD 34908UI00

Patients

Seq Age Sex Outcome Treatment
1 ARCHITECT I2000SR ANALYZER 3M74-02| SN (B)(4)| ARCHITECT I2000SR ANALYZER 3M74-02| SN (B)(4)