FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION, DF4 CONNECTOR

MDR report key: 3781842 · Received April 30, 2014

Report

Report Number
2938836-2014-10028
Event Type
Injury
Date Received
April 30, 2014
Date of Event
February 6, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. EVALUATION DESCRIPTION: THE DAMAGE FOUND WAS SUSTAINED DURING THE SURGICAL PROCEDURE. THE LEAD WAS OTHERWISE NORMAL.

Description of Event or Problem · 1

IT WAS REPORTED THAT A MERLIN.NET TRANSMISSION REVEALED INTERMITTENT VENTRICULAR UNDERSENSING. ALTHOUGH PROGRAMMING CHANGES WERE MADE TO RESOLVE THE UNDERSENSING ISSUE, THE PHYSICIAN ELECTED TO EXPLANT AND REPLACE THE LEAD DUE TO HISTORICALLY LOW R-WAVES. THE PATIENT WAS FINE AFTER THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
259106 DURATA STS OPTIM ACTIVE FIXATION, DF4 CONNECTOR DEFIBRILLATION LEAD NVY ST. JUDE MEDICAL, INC., CRMD 7120Q/65 NA

Patients

Seq Age Sex Outcome Treatment
1 53 YR Required Intervention