FDA Adverse Event
Injury
Summary report: N
DURATA STS OPTIM ACTIVE FIXATION
MDR report key: 3781816
·
Received April 30, 2014
Report
- Report Number
- 2938836-2014-09980
- Event Type
- Injury
- Date Received
- April 30, 2014
- Date of Event
- February 4, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
A PARTIAL LEAD WITH THE CONNECTOR PIN MEASURING 13.4CM WAS RETURNED FOR ANALYSIS. THE DAMAGE FOUND WAS SUSTAINED DURING THE SURGICAL PROCEDURE. THE PORTION OF THE LEAD THAT WAS RETURNED WAS OTHERWISE NORMAL. WITHOUT RETURN OF THE ENTIRE LEAD, A COMPLETE ANALYSIS COULD NOT BE PERFORMED.
Description of Event or Problem · 1
IT WAS REPORTED HIGH, OUT OF RANGE, PACING LEAD IMPEDANCE AND HIGH THRESHOLD WERE OBSERVED. THE LEAD WAS CAPPED AND REPLACED. THE PATIENT CONDITION WAS NORMAL AFTER THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 260795 | DURATA STS OPTIM ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST. JUDE MEDICAL, INC., CRMD | 7121/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR | Required Intervention |