FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION

MDR report key: 3781800 · Received April 30, 2014

Report

Report Number
2938836-2014-10020
Event Type
Injury
Date Received
April 30, 2014
Date of Event
March 25, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A PARTIAL LEAD WITH THE CONNECTOR PIN MEASURING 3.6CM WAS RETURNED FOR ANALYSIS. THE DAMAGE FOUND WAS SUSTAINED DURING THE SURGICAL PROCEDURE. THE PORTION OF THE LEAD THAT WAS RETURNED WAS OTHERWISE NORMAL. WITHOUT RETURN OF THE ENTIRE LEAD, A COMPLETE ANALYSIS COULD NOT BE PERFORMED.

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

THE SENSE/PACE PORTION WAS RETURNED FOR ANALYSIS.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ASYMPTOMATIC PATIENT PRESENTED IN THE OPERATING ROOM FOR NORMAL ERI DEVICE CHANGE OUT. HIGH, OUT OF RANGE, PACING LEAD IMPEDANCE AND HIGH CAPTURE THRESHOLD WERE OBSERVED. THE PACE/SENSE PORTION OF THE LEAD WAS CAPPED AND REPLACED. THE DEFIB PORTION REMAINS ACTIVE. THE PATIENT CONDITION WAS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
260331 DURATA STS OPTIM ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST. JUDE MEDICAL, INC., CRMD 7121/65 NA

Patients

Seq Age Sex Outcome Treatment
1 81 YR Required Intervention (B)(4)