FDA Adverse Event
Injury
Summary report: N
DURATA STS OPTIM ACTIVE FIXATION
MDR report key: 3781800
·
Received April 30, 2014
Report
- Report Number
- 2938836-2014-10020
- Event Type
- Injury
- Date Received
- April 30, 2014
- Date of Event
- March 25, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
A PARTIAL LEAD WITH THE CONNECTOR PIN MEASURING 3.6CM WAS RETURNED FOR ANALYSIS. THE DAMAGE FOUND WAS SUSTAINED DURING THE SURGICAL PROCEDURE. THE PORTION OF THE LEAD THAT WAS RETURNED WAS OTHERWISE NORMAL. WITHOUT RETURN OF THE ENTIRE LEAD, A COMPLETE ANALYSIS COULD NOT BE PERFORMED.
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.
Description of Event or Problem · 1
THE SENSE/PACE PORTION WAS RETURNED FOR ANALYSIS.
Description of Event or Problem · 1
IT WAS REPORTED THAT AN ASYMPTOMATIC PATIENT PRESENTED IN THE OPERATING ROOM FOR NORMAL ERI DEVICE CHANGE OUT. HIGH, OUT OF RANGE, PACING LEAD IMPEDANCE AND HIGH CAPTURE THRESHOLD WERE OBSERVED. THE PACE/SENSE PORTION OF THE LEAD WAS CAPPED AND REPLACED. THE DEFIB PORTION REMAINS ACTIVE. THE PATIENT CONDITION WAS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 260331 | DURATA STS OPTIM ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST. JUDE MEDICAL, INC., CRMD | 7121/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 81 YR | Required Intervention | (B)(4) |