FDA Adverse Event Injury Summary report: N

FORTIFY DR, DF-4 CONNECTOR

MDR report key: 3781744 · Received April 30, 2014

Report

Report Number
2938836-2014-09885
Event Type
Injury
Date Received
April 30, 2014
Date of Event
February 28, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
LWS
PMA / PMN Number
NA
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DO
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE HV THERAPY WHEN THE ATP THERAPY ACCELERATED THE RHYTHM TO THE VF ZONE. THE DEVICE WAS REPROGRAMMED AND THE PATIENT HAD A SUCCESSFUL VT ABLATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
260109 FORTIFY DR, DF-4 CONNECTOR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS ST. JUDE MEDICAL, INC., CRMD CD2233-40Q NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention