FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 3781711 · Received April 30, 2014

Report

Report Number
1416980-2014-14012
Event Type
Malfunction
Date Received
April 30, 2014
Date of Event
April 9, 2014
Report Date
April 9, 2014
Manufacturer
SHARP CORPORATION OSAKA JAPAN
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. THIS IS AN ANCILLARY SERVICE EVENT. THE CAUSE OF THE F-74 ALARM WAS DETERMINED TO BE AN INOPERATIVE CENTRAL PROCESSING UNIT (CPU) BOARD. TO CORRECT THE CONDITION, THE DEVICE WAS REMOVED FROM SERVICE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING PRODUCT SERVICE BY A SERVICE TECHNICIAN, A FLOGARD INFUSION PUMP WAS FOUND TO HAVE PRESENTED AN F-74 ALARM. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
259614 FLOGARD PUMP, INFUSION FRN SHARP CORPORATION OSAKA JAPAN

Patients

Seq Age Sex Outcome Treatment
1