FDA Adverse Event Malfunction Summary report: N

ULTRACISION HARMONIC ACE

MDR report key: 3781685 · Received April 30, 2014

Report

Report Number
3005075853-2014-02939
Event Type
Malfunction
Date Received
April 30, 2014
Date of Event
April 18, 2014
Report Date
April 22, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
LFL
PMA / PMN Number
K042777
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: THE DEVICE WAS RETURNED WITH THE DISTAL TIP OF THE BLADE BROKEN OFF AND NOT RETURNED. THE REMAINING BLADE PORTION WAS SCRATCHED WITH C-STAPLE MARKS. THIS BLADE TIP PORTION MAY HAVE BROKEN OFF OF THE DEVICE DURING TRANSPORT TO OUR ANALYSIS SITE. THE REPORTED COMPLAINT WAS FOR THE TISSUE PAD DETACHED. UPON VISUAL INSPECTION THE TISSUE PAD WAS FOUND TO BE PROPERLY ATTACHED TO THE CLAMP ARM; HOWEVER IT WAS FOUND DAMAGED WITH C-STAPLE MARKS. THE DEVICE WAS FUNCTIONALLY TESTED WITH A GENERATOR. DURING FUNCTIONAL TESTING ON GEN11 AN ALERT SCREEN WAS DISPLAYED. A PROBABLE CAUSE OF AN ALERT SCREEN IS BLADE DAMAGE. PROBABLE CAUSES OF BLADE DAMAGE, INCLUDING BREAKAGE, ARE EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE, BLADE CONTACT WITH OTHER DEVICES, STAPLES OR CLIPS DURING THE PROCEDURE OR USING ANY MEANS OTHER THAN THE BLADE WRENCH TO ATTACH OR DETACH THE BLADE. ONCE MINOR BLADE DAMAGE HAS OCCURRED, SUBSEQUENT ACTIVATIONS MAY INCREASE DAMAGE SEVERITY AND RESULT IN ALERT SCREENS, SUCH AS ¿TIGHTEN ASSEMBLY¿, ¿BLADE ERROR DETECTED¿ OR "RELAXED PRESSURE ON BLADE" FOLLOWED BY A ¿REPLACE INSTRUMENT¿ SCREEN LATER IN THE PROCEDURE. PROBABLE CAUSES OF THE TISSUE PAD DAMAGE ARE EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE, BLADE CONTACT WITH OTHER DEVICES, STAPLES OR CLIPS DURING THE PROCEDURE OR USING ANY MEANS OTHER THAN THE BLADE WRENCH TO ATTACH OR DETACH THE BLADE.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. NO DEVICE RECEIVED FOR ANALYSIS AT TIME OF SUBMISSION OF 3500A. WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A BYPASS PROCEDURE THE DEVICE THE TISSUE PAD DETACHED AND WAS INSIDE THE PATIENT, THE PART WAS RETRIEVED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE CASE. NO PATIENT CONSEQUENCES. ONE DEVICE WILL BE RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
259956 ULTRACISION HARMONIC ACE INSTRUMENT, ULTRASONIC SURGICAL LFL ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR AND HANDPIECE