FDA Adverse Event Malfunction Summary report: N

5.0CM SHORT ATTACHMENT

MDR report key: 3781630 · Received April 30, 2014

Report

Report Number
1045834-2014-11953
Event Type
Malfunction
Date Received
April 30, 2014
Date of Event
April 10, 2014
Report Date
April 17, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
PK011444
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. (B)(4) EVALUATED THE DEVICE AND THE REPORTED CONDITION WAS CONFIRMED. A FUNCTIONAL ASSESSMENT WAS PERFORMED ON THE DEVICE WHICH FOUND THAT THE HEATING UP BEYOND REQUIRED SPECIFICATIONS WAS CAUSED BY WORN BEARINGS. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO NORMAL WEAR OVER TIME. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO SURGERY, IT WAS DISCOVERED THAT THE ATTACHMENT DEVICE WAS ¿HEATING UP¿. THERE WERE NO DELAYS TO THE SURGICAL PROCEDURE AS AN IDENTICAL SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
259900 5.0CM SHORT ATTACHMENT MOTOR, DRILL, ELECTRIC - ATTACHMENT HBC DEPUY SYNTHES POWER TOOLS NA

Patients

Seq Age Sex Outcome Treatment
1