5.0CM SHORT ATTACHMENT
Report
- Report Number
- 1045834-2014-11953
- Event Type
- Malfunction
- Date Received
- April 30, 2014
- Date of Event
- April 10, 2014
- Report Date
- April 17, 2014
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- HBC
- PMA / PMN Number
- PK011444
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NURSE
Narratives
THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. (B)(4) EVALUATED THE DEVICE AND THE REPORTED CONDITION WAS CONFIRMED. A FUNCTIONAL ASSESSMENT WAS PERFORMED ON THE DEVICE WHICH FOUND THAT THE HEATING UP BEYOND REQUIRED SPECIFICATIONS WAS CAUSED BY WORN BEARINGS. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO NORMAL WEAR OVER TIME. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
IT WAS REPORTED THAT PRIOR TO SURGERY, IT WAS DISCOVERED THAT THE ATTACHMENT DEVICE WAS ¿HEATING UP¿. THERE WERE NO DELAYS TO THE SURGICAL PROCEDURE AS AN IDENTICAL SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.