PATELLA CLAMP
Report
- Report Number
- 0002249697-2014-01575
- Event Type
- Malfunction
- Date Received
- April 30, 2014
- Date of Event
- April 7, 2014
- Report Date
- April 7, 2014
- Manufacturer
- STRYKER ORTHOPAEDICS-MAHWAH
- Product Code
- JWH
- PMA / PMN Number
- K123486
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
Narratives
AN EVENT REGARDING DEGRADED HANDLES INVOLVING A TRIATHLON PATELLAR CLAMP WAS REPORTED. THE EVENT WAS CONFIRMED. METHOD AND RESULTS: DEVICE EVALUATION AND RESULTS: VISUAL ANALYSIS CONFIRMED THE REPORTED EVENT. MEDICAL RECORDS RECEIVED AND EVALUATION: NOT PERFORMED BECAUSE IT IS BELIEVED PATIENT FACTORS DID NOT CONTRIBUTE TO THE REPORTED EVENT. DEVICE HISTORY REVIEW INDICATED ALL DEVICES ACCEPTED INTO STOCK MET SPECIFICATION. COMPLAINT HISTORY REVIEW INDICATED THERE HAVE BEEN NO OTHER EVENTS ASSOCIATE WITH THE REPORTED LOT. CONCLUSION: CAPA WAS INITIATED BECAUSE A SERIES OF COMPLAINTS WERE RECEIVED FOR TRIATHLON INSTRUMENTATION WHERE GREEN SANTOPRENE OVER-MOLD HANDLES HAVE BEEN REPORTED TO BE DEGRADING, BECOMING STICKY AND UNSTABLE. THIS EVENT WAS DETERMINED TO BE UNDER THE SCOPE OF CAPA. CHEMICAL AND CYTOTOXICITY TESTS SHOWED THE DEGRADED SANTOPRENE IS NON-TOXIC. THE ABILITY TO CLEAN AND STERILIZE THE DEGRADED INSTRUMENTS HAS NOT BEEN COMPROMISED.
A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.
IT WAS REPORTED THAT WHILE CHECKING INSTRUMENTS, I NOTICED THE RUBBER HANDLES ON THE PATELLA CLAMP, AND THE PATELLA CALIPER WERE STICKY AND THE RUBBER APPEARS TO BE UNSTABLE AND EASILY SCRAPED WITH A FINGER NAIL.
IT WAS REPORTED THAT WHILE CHECKING INSTRUMENTS I NOTICED THE RUBBER HANDLES ON THE PATELLA CLAMP, AND THE PATELLA CALIPER WERE STICKY AND THE RUBBER APPEARS TO BE UNSTABLE AND EASILY SCRAPED WITH A FINGER NAIL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 260151 | PATELLA CLAMP | INSTRUMENT | JWH | STRYKER ORTHOPAEDICS-MAHWAH | C3S18 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |