FDA Adverse Event Injury Summary report: N

132 SIZE 8 SECUR-FIT ADVANCED STEM

MDR report key: 3781259 · Received April 30, 2014

Report

Report Number
0002249697-2014-01621
Event Type
Injury
Date Received
April 30, 2014
Date of Event
April 8, 2014
Report Date
April 9, 2014
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
LWJ
PMA / PMN Number
K122853
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AN EVALUATION OF THE DEVICE CANNOT BE PERFORMED AS THE DEVICE REMAINED IMPLANTED IN THE PATIENT AND WAS NOT RETURNED TO THE MANUFACTURER. ADDITIONAL INFORMATION HAS BEEN REQUESTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE IT WILL BE REPORTED IN A SUPPLEMENTAL REPORT UPON COMPLETION OF THE INVESTIGATION. NOT RETURNED TO THE MANUFACTURER.

Additional Manufacturer Narrative · 1

THE PATIENT IS (B)(6) IN HEIGHT. AN EVENT REGARDING STEM SUBSIDENCE INVOLVING A SECURFIT STEM WAS REPORTED. THE EVENT WAS CONFIRMED. METHOD AND RESULTS: MEDICAL RECORDS RECEIVED AND EVALUATION: THE PROVIDED X-RAYS WERE SUBMITTED TO A CONSULTING CLINICIAN WHO DEEMED IT INSUFFICIENT FOR A MEDICAL REVIEW. STEM SUBSIDENCE CONFIRMED ON X-RAY HOWEVER OPERATIVE REPORTS AND CLINICAL HISTORY ARE NEEDED TO FURTHER INVESTIGATE. CONCLUSIONS: A MEDICAL REVIEW WAS PERFORMED AND CONFIRMED THAT THE STEM SUBSIDED HOWEVER THE EXACT CAUSE OF THE EVENT COULD NOT BE DETERMINED. ADDITIONAL INFORMATION, INCLUDING OPERATIVE REPORTS, PROGRESS NOTES, CLINICAL HISTORY AND RETURN OF THE DEVICE ARE NEEDED TO FULLY INVESTIGATE THE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DOCTOR OBSERVED FEMORAL STEM SUBSIDENCE OF 6MM ON THE 6 WEEK X-RAYS FOR A PATIENT PARTICIPATING IN THE (B)(4) CLINICAL STUDY.

Description of Event or Problem · 1

IT WAS REPORTED THAT DOCTOR OBSERVED FEMORAL STEM SUBSIDENCE OF 6MM ON THE 6 WEEK X-RAYS FOR A PATIENT PARTICIPATING IN THE SECUR-FIT ADVANCED POST-MARKET CLINICAL STUDY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
259272 132 SIZE 8 SECUR-FIT ADVANCED STEM IMPLANT LWJ STRYKER ORTHOPAEDICS-MAHWAH MMO52K

Patients

Seq Age Sex Outcome Treatment
1 71 YR Other