FDA Adverse Event
Malfunction
Summary report: N
SPECTRUM INFUSION PUMP
MDR report key: 3780628
·
Received December 3, 2013
Report
- Report Number
- 1314492-2013-19222
- Event Type
- Malfunction
- Date Received
- December 3, 2013
- Date of Event
- October 1, 2013
- Report Date
- November 5, 2013
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE DEVICE WAS RETURNED AN EVALUATED BY BAXTER. THIS COMPLAINT WAS OPENED DURING THE INVESTIGATION OF COMPLAINT (B)(4). THE IDENTIFIED "DOOR JAMMED" ALARMS WERE CONFIRMED THROUGH THE EVENT HISTORY LOG REVIEW. THE CAUSE WAS UNDETERMINED. IF ANY ADDITIONAL INFORMATION IS RECEIVED A FOLLOW UP WILL BE SENT. THE INPUT/OUTPUT PRINTED CIRCUIT BOARD ASSEMBLY WAS REPLACED.
Description of Event or Problem · 1
DURING THE EVALUATION OF A RETURNED SPECTRUM INFUSION PUMP, 2 OCCURRENCES OF "DOOR JAMMED" ALARMS WERE IDENTIFIED IN THE EVENT HISTORY LOG. ANY PATIENT INJURY OR MEDICAL INTERVENTION IS UNKNOWN SINCE THESE ITEMS WERE FOUND DURING EVALUATION OF THE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 626839 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE CORPORATION | 35700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |