FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 3780200 · Received January 3, 2014

Report

Report Number
1314492-2014-00504
Event Type
Malfunction
Date Received
January 3, 2014
Date of Event
November 1, 2013
Report Date
December 6, 2013
Manufacturer
BAXTER HEALTHCARE CORP
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY BAXTER. THIS COMPLAINT WAS OPENED DURING THE INVESTIGATION OF COMPLAINT (B)(4). THE "DOOR JAMMED" ALARMS WERE CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS UNDETERMINED. IF ANY ADDITIONAL INFO IS RECEIVED A FOLLOW UP WILL BE SENT. THE DOOR LATCH HOOK, AND INPUT/OUTPUT PRINTED CIRCUIT BOARD WERE REPLACED.

Description of Event or Problem · 1

DURING THE EVAL OF A RETURNED SPECTRUM INFUSION PUMP, 6 OCCURRENCES OF "DOOR JAMMED" ALARMS WERE IDENTIFIED IN THE EVENT HISTORY LOG. ANY PT INVOLVEMENT, INJURY OR MEDICAL INTERVENTION IS UNK SINCE THESE ITEMS WERE FOUND DURING EVAL OF THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
4921 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE CORP 35700

Patients

Seq Age Sex Outcome Treatment
1