FDA Adverse Event
Malfunction
Summary report: N
SPECTRUM INFUSION PUMP
MDR report key: 3780137
·
Received January 3, 2014
Report
- Report Number
- 1314492-2014-00438
- Event Type
- Malfunction
- Date Received
- January 3, 2014
- Date of Event
- November 1, 2013
- Report Date
- December 6, 2013
- Manufacturer
- BAXTER HEALTHCARE CORP
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
MFR REF NO.: (B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY BAXTER. THIS COMPLAINT WAS OPENED DURING THE INVESTIGATION OF COMPLAINT (B)(4). THE IDENTIFIED DOWNSTREAM OCCLUSION ALARMS WERE CONFIRMED THROUGH THE EVENT HISTORY LOG REVIEW. THE CAUSE WAS UNDETERMINED. IF ANY ADDITIONAL INFORMATION IS RECEIVED A FOLLOW UP WILL BE SENT. THE INPUT/OUTPUT PRINTED CIRCUIT BOARD, DOWNSTREAM TUBING GUIDE, AND FORCE SENSOR GASKET WERE REPLACED.
Description of Event or Problem · 1
DURING THE EVALUATION OF A RETURNED SPECTRUM INFUSION PUMP, 188 OCCURRENCES OF DOWNSTREAM OCCLUSION ALARMS WERE IDENTIFIED IN THE EVENT HISTORY LOG. ANY PATIENT INVOLVEMENT, INJURY OR MEDICAL INTERVENTION IS UNKNOWN SINCE THESE ITEMS WERE FOUND DURING EVALUATION OF THE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 4669 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE CORP | 35700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |