FDA Adverse Event Other Summary report: N

BARI AIR BED

MDR report key: 377786 · Received February 18, 2002

Report

Report Number
MW1024103
Event Type
Other
Date Received
February 18, 2002
Date of Event
June 12, 2001
Report Date
February 15, 2002
Manufacturer
KCI
Product Code
FNL
Report Source
Voluntary report
Reporter Location
VA, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

PT WAS BEING MOVED WHEN A WHEEL BROKE ON THE BARI AIR BED. COULDN'T MOVE BED OFF THE ELEVATOR. STAFF GOT HYDRAULIC LIFT TO SAFELY MOVE PT OUT OF THE ELEVATOR. PT HAD A PATHOLOGIC FRACTURE AND COULDN'T BE MOVED OUT OF THE BED. BED HAD TO BE REPLACED WHILE PT WAS IN IT.

Description of Event or Problem · 1

ADD'L INFO REC'D FROM MFR 4/19/02: THE BARIAIR THERAPY SYSTEM BED WAS PLACED ON THE MARKET IN 10/97. EXAMINATION OF THE BROKEN CASTERS REVEALED THE BREAK OCCURRED AT THE ALUMINUM HOUSING THAT ATTACHED THE CASTER TO THE BED CASTER STEM. IT WAS DETERMINED THAT THE ALUMINUM HOUSING HAD WEAKENED DUE TO NUMEROUS TRANSPORTING OVER THE PAST SEVERAL YEARS, TO AND FROM THE HOSPS, WITH REPEATED IMPACTS OVER CURBS AND DOORWAY THRESHOLDS AND IN AND OUT OF ELEVATORS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BARI AIR BED BARIATRIC PT BED FNL KCI UNK *

Patients

Seq Age Sex Outcome Treatment
1 58 YR Other