FDA Adverse Event Malfunction Summary report: N

PROGRASP FORCEPS INSTRUMENT

MDR report key: 3777461 · Received April 29, 2014

Report

Report Number
2955842-2014-02622
Event Type
Malfunction
Date Received
April 29, 2014
Date of Event
April 2, 2014
Report Date
April 11, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS INVESTIGATION WAS UNABLE TO CONFIRM THE REPORTED COMPLAINT. VISUAL INSPECTION FOUND NO DAMAGED, FRAYED, BROKEN, OR LOOSE CABLES. FAILURE ANALYSIS INVESTIGATION FOUND THAT THE DISTAL END OF THE MAIN TUBE HAD VARIOUS SCRATCH MARKS EXHIBITING LIGHT MATERIAL REMOVAL. THE SCRATCHES WERE APPROXIMATELY .032 - .083 IN LENGTH AND NOT AXIALLY ALIGNED WITH THE TUBE. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE TUBE ABRASIONS WITH MATERIAL REMOVED FOUND DURING FAILURE ANALYSIS INVESTIGATION, COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT, IF THE MALFUNCTION WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE REPROCESSING OF A DA VINCI PROGRASP FORCEPS INSTRUMENT, A FRAYED WIRE WAS IDENTIFIED. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO A PATIENT, AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
257739 PROGRASP FORCEPS INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420093-08 M11121212 855

Patients

Seq Age Sex Outcome Treatment
1