NXSTAGE SYSTEM ONE
Report
- Report Number
- 3003464075-2014-00007
- Event Type
- Death
- Date Received
- April 18, 2014
- Date of Event
- February 27, 2014
- Report Date
- February 27, 2014
- Manufacturer
- NXSTAGE MEDICAL, INC.
- Product Code
- KDI
- PMA / PMN Number
- K050525
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NURSE
Narratives
THE EXACT CAUSE OF THE PT'S DEATH CANNOT BE DETERMINED. THE EXTRACORPOREAL CARTRIDGE WAS DISPOSED OF AND THE LOT NUMBER WAS NOT PROVIDED. THE CARTRIDGE IFU STATES: DO NOT USE THE CARTRIDGE FOR MORE THAN 72 HOURS OR 864 LITERS OF BLOOD PROCESSED. ADDITIONALLY PER THE USER'S GUIDE WHEN THE CAUTION 70 OCCURS, THE CARTRIDGE SHOULD BE CHANGED OUT. NO FURTHER INVESTIGATION CAN BE PERFORMED. NXTAGE MEDICAL CONSIDERS THIS REPORT CLOSED. NO ADDITIONAL INFORMATION WILL BE PROVIDED.
PT WITH COPD, HYPERTENSION, TYPE II DIABETES, AND ACUTE KIDNEY INJURY WAS ADMITTED TO ICU FROM THE EMERGENCY ROOM ON (B)(6) 2014. PT STARTED ON CVVHD ON (B)(6) 2014 WITHOUT INCIDENT. CARTRIDGE CHANGE OCCURRED ON (B)(6) 2014. ON (B)(6) 2014 AT 6:41, A CAUTION 70 OCCURRED INDICATING THE CARTRIDGE LIFE EXPECTANCY HAD BEEN EXCEEDED. TREATMENT WAS ENDED AT 7:33 ON (B)(6) 2014 AND THE PT'S BLOOD WAS RETURNED. DURING THE 52 MINUTES OF THE CAUTION 70, THE BLOOD PUMP CONTINUED TO RUN HOWEVER FILTERING OF THE PT'S BLOOD PUMP CONTINUED TO RUN; HOWEVER, FILTERING OF THE PT'S BLOOD HAD DISCONTINUED. THE PT RAPIDLY BECAME HYPOTENSIVE, MOTTLED AND CODED. THE BLOOD IN THE DIALYSATE LINES WAS NOTED AS SEPARATED. THE PT EXPIRED AFTER TREATMENT WAS DISCONTINUED (OTHERWISE UNSPECIFIED). THE PHYSICIAN BELIEVES THIS EVENT TO BE A COMPLEMENT MEDIATED REACTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 239032 | NXSTAGE SYSTEM ONE | HIGH PERMEABILITY HEMODIALYSIS SYSTEM | KDI | NXSTAGE MEDICAL, INC. | CAR-500 | NI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR | Death |