FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3772184 · Received January 2, 2014

Report

Report Number
1720753-2014-00009
Event Type
Malfunction
Date Received
January 2, 2014
Date of Event
December 11, 2013
Report Date
January 2, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS(SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE 5VDC POWER SUPPLY, GIB PCB ASSEMBLY, PASSIVE BACKPLANE AND THE MAINFRAME BATTERY PACKS WERE EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED CHARGING ERRORS UPON BOOT UP AND FAILED TO ALLOW FLUOROSCOPIC EXPOSURES. NO PT SERIOUS INJURY OR DEATH WAS REPORTED RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2819 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS(SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1