FDA Adverse Event Malfunction Summary report: N

STENOSCOP

MDR report key: 3772138 · Received January 2, 2014

Report

Report Number
1720753-2014-00001
Event Type
Malfunction
Date Received
January 2, 2014
Date of Event
December 10, 2013
Report Date
January 2, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE MDA CPU AND ASSOCIATED MEMORY CONNECTIONS WERE EVALUATED AND REASSEMBLED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM FAILED TO BOOT TO A USABLE STATE AND EXHIBITED A NON-RECOVERABLE LOSS OF FUNCTIONALITY. NO PATIENT SERIOUS INJURY OR DEATH WAS REPORTED RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2687 STENOSCOP FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) STENOSCOP

Patients

Seq Age Sex Outcome Treatment
1