FDA Adverse Event Other Summary report: N

SYSMEX CA-1500 ANALYZER

MDR report key: 3772126 · Received April 11, 2014

Report

Report Number
3009711478-2014-00009
Event Type
Other
Date Received
April 11, 2014
Report Date
March 27, 2014
Manufacturer
SYSMEX CORPORATION
Product Code
GKP
PMA / PMN Number
K993299
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
DE, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE SAMPLE PLASMA WAS CLEAR, FREE OF CLOTS AND LESS THAN 4 HRS OLD. THE OPERATOR DISPENSES ALL SAMPLES INTO 4ML SAMPLE CUPS FOR ANALYSIS ON THE ANALYZER. THE OPERATOR CONFIRMED ANALYZER PERFORMANCE WITH QUALITY CONTROL AFTER THE EVENT OCCURRED. THE OPERATOR ALSO CONFIRMED ANALYZER REPRODUCIBILITY BY PERFORMING A PRECISION CHECK. BOTH QC AND PRECISION PASSED WITHIN LABORATORY'S SPECIFICATION. THE OPERATOR DECLINED FIELD SERVICE ENGINEER (FSE) DISPATCH. RECENT SERVICE ON (B)(6) 2014 WAS PERFORMED AND UNIT WAS OPERATIONAL AT THAT TIME. REVIEW OF THE SAMPLE RESULT FOR INITIAL PT WAS 19.0 SECONDS WITH REPEAT ANALYSIS RESULT OF 9.8 SECONDS. ROOT CAUSE OF THE SIGNIFICANT DIFFERENCE BETWEEN ANALYSES WAS NOT DETERMINED AT THE TIME OF THIS ASSESSMENT. SAMPLE HANDLING AND ANALYZER PERFORMANCE COULD NOT BE RULED OUT. THIS INCIDENT IS REPORTED AS ROOT CAUSE OF THE OCCURRENCE COULD NOT BE CONFIRMED.

Description of Event or Problem · 1

THE USER OF A SYSMEX CA-1500 S/N (B)(4) REPORTED TO SIEMENS HEALTHCARE DIAGNOSTICS ON (B)(4) 2014 REGARDING ERRONEOUS RESULTS GENERATED FOR INTERNATIONAL NORMALIZED RATIO (INR). THE USER ANALYZED A SAMPLE FROM AN ER PT SUSPECTED OF SUFFERING A STROKE. THE INITIAL RESULT INR=1.82, PROTHROMBIN TIME (PT) WAS 19.0 SECONDS, RESULTS REPORTED. THE ER DOCTOR QUESTIONED THE RESULT BECAUSE THE LABORATORY RESULTS DID NOT MEET CRITERIA TO ADMINISTER TISSUE PLASMINOGEN ACTIVATOR (TPA); INR <1.7. THE DR REQUESTED A REDRAW. THE REDRAW RESULT INR=1.03. THE OPERATOR ANALYZED QC AND PARAMETERS WERE WITHIN THE LAB'S SPECIFIED RANGE. THE OPERATOR THEN PERFORMED PRECISION WHICH PASSED. REPEAT ANALYSIS ON THE INITIAL DRAW WAS PT 9.8 SECONDS, INR 0.97. THE OPERATOR SEN THE INITIAL DRAW TO THE MAIN LAB FOR REPEAT ANALYSIS ON A BEHRING COAGULATION SYSTEM (BCS) ANALYZER, MODEL UNK, AND GENERATED AN INR OF 1.02. REPEAT ANALYSIS AND NEW SAMPLE COLLECTION DELAYS POTENTIAL TPA TREATMENT FOR PTS WHO PRESENT WITH SYMPTOMS OF STROKE. THIS PT WAS EXCLUDED AS A CANDIDATE FOR TPA FOR OTHER CRITERIA. THE OCCURRENCE WAS REPORTED TO SYSMEX AMERICA, INC ON (B)(6) 2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
221584 SYSMEX CA-1500 ANALYZER AUTOMATED BLOOD COAGULATION ANALYZER GKP SYSMEX CORPORATION CA-1500

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention