FDA Adverse Event Injury Summary report: N

M2A-MAGNUM PF CUP 48ODX42ID

MDR report key: 3772114 · Received April 25, 2014

Report

Report Number
0001825034-2014-03267
Event Type
Injury
Date Received
April 25, 2014
Date of Event
May 28, 2013
Report Date
June 3, 2014
Manufacturer
BIOMET ORTHOPEDICS
Product Code
KWA
PMA / PMN Number
PK042037
Removal / Correction Number
N/A
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
ATTORNEY

Narratives

Additional Manufacturer Narrative · 1

THIS FOLLOW-UP REPORT IS BEING FILED TO RELAY ADDITIONAL INFORMATION, WHICH WAS UNKNOWN AT THE TIME OF THE INITIAL MEDWATCH.

Additional Manufacturer Narrative · 1

CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER POSSIBLE ADVERSE EFFECTS, NUMBER 1 STATES, ¿MATERIAL SENSITIVITY REACTIONS.¿ NUMBER 6 STATES, ¿INADEQUATE RANGE OF MOTION DUE TO IMPROPER SELECTION OR POSITIONING OF COMPONENTS.¿ NUMBER 14 STATES, ¿INTRAOPERATIVE OR POSTOPERATIVE BONE FRACTURE AND/OR POSTOPERATIVE PAIN.¿ NUMBER 15 STATES, ¿ELEVATED METAL ION LEVELS HAVE BEEN REPORTED WITH METAL-ON-METAL ARTICULATING SURFACES.¿ THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF¿S COMPLAINT AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED. THIS REPORT IS NUMBER 2 OF 2 MDRS FILED FOR THE SAME EVENT (REFERENCE 1825034-2014-03266 & 03267).

Description of Event or Problem · 1

LEGAL COUNSEL FOR PATIENT REPORTED PATIENT UNDERWENT A LEFT TOTAL HIP ARTHROPLASTY ON (B)(6) 2008. PATIENT'S LEGAL COUNSEL FURTHER REPORTED THAT A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2013., DUE TO PATIENT ALLEGATIONS OF PAIN, DISCOMFORT, INFLAMMATION, LOSS OF RANGE OF MOTION, DYSFUNCTION, METAL POISONING, METALLOSIS AND ELEVATED METAL ION LEVELS. A REVIEW OF THE INVOICE HISTORY CONFIRMED THE SURGERY DATES AND THAT THE ACETABULAR CUP, MODULAR HEAD AND TAPER ADAPTER WERE REMOVED AND REPLACED DURING THE REVISION PROCEDURE. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF¿S COMPLAINT AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED.

Description of Event or Problem · 1

LEGAL COUNSEL FOR PATIENT REPORTED PATIENT UNDERWENT A LEFT TOTAL HIP ARTHROPLASTY ON (B)(6) 2008. PATIENT'S LEGAL COUNSEL FURTHER REPORTED THAT A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2013 DUE TO PATIENT ALLEGATIONS OF PAIN, DISCOMFORT, INFLAMMATION, LOSS OF RANGE OF MOTION, DYSFUNCTION, METAL POISONING, METALLOSIS AND ELEVATED METAL ION LEVELS. A REVIEW OF THE INVOICE HISTORY CONFIRMED THE SURGERY DATES AND THAT THE ACETABULAR CUP, MODULAR HEAD AND TAPER ADAPTER WERE REMOVED AND REPLACED DURING THE REVISION PROCEDURE. ADDITIONAL INFORMATION RECEIVED IN PATIENT MEDICAL RECORDS INDICATES THAT LEFT HIP REVISION PROCEDURE PERFORMED ON (B)(6) 2013 WAS DUE TO PAIN AND DIFFICULTY AMBULATING. THE PATIENT铠OPERATIVE REPORT NOTED METALLIC FLUID, METALLOSIS, DAMAGE TO SYNOVIAL TISSUE, EROSION OF EXTERNAL ROTATORS/SURROUNDING TISSUE AND A DISPLACED ACETABULAR CUP. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF'S COMPLAINT AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
251532 M2A-MAGNUM PF CUP 48ODX42ID PROSTHESIS, HIP KWA BIOMET ORTHOPEDICS N/A 048510

Patients

Seq Age Sex Outcome Treatment
1 49 YR Hospitalization| R