FDA Adverse Event Malfunction Summary report: N

8800

MDR report key: 3772087 · Received January 2, 2014

Report

Report Number
1720753-2014-00046
Event Type
Malfunction
Date Received
January 2, 2014
Date of Event
December 16, 2013
Report Date
January 2, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE CONNECTOR CABLE BETWEEN THE 5V POWER SUPPLY AND THE DISPLAY ADAPTER PCB WAS EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS A COMMUNICATION ERROR MESSAGE. THIS ERROR MAY RESULT IN A SYSTEM LOCK UP, NO BOOT, OR SHUT DOWN. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.