ANIMAS INSULIN CARTRIDGE
Report
- Report Number
- 2531779-2014-11364
- Event Type
- Malfunction
- Date Received
- April 25, 2014
- Report Date
- April 11, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K032257
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
THE CARTRIDGE HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE CARTRIDGE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE (OCCLUSION AND LOSS OF PRIME), CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION.
(B)(4)-DEVICE EVALUATION: THE CARTRIDGE HAS NOT BEEN RETURNED TO ANIMAS. A RETAIN SAMPLE FROM THE SAME LOT OF CARTRIDGE WAS TESTED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: A VISUAL INSPECTION OF THE RETAIN CARTRIDGE WAS PERFORMED. NO DAMAGE OR DEFECTS WERE NOTED. A FILL TEST, A FORCE TEST AND A LEAK TEST WERE PERFORMED ON THE RETAIN SAMPLE AND IT MET PASSING CRITERIA. THE CARTRIDGE FORCE READINGS WERE CONFIRMED TO BE WITHIN SPECIFICATIONS; NO FAILURES RELATED TO THE LOSS OF PRIME COMPLAINT WERE NOTED. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE (OCCLUSION AND LOSS OF PRIME), CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION. A REVIEW OF INCOMING INSPECTION RECORD INDICATED THAT THE LOT MET RELEASE CRITERIA.
(B)(4)-DEVICE EVALUATION: THE CARTRIDGE WAS RETURNED TO ANIMAS AND TESTED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: A VISUAL INSPECTION OF THE RETURNED CARTRIDGE WAS PERFORMED. NO DAMAGE OR DEFECTS WERE NOTED. A FORCE TEST AND A FILL TEST WERE PERFORMED ON THE RETURNED PRODUCT AND IT MET PASSING CRITERIA. A REVIEW OF INCOMING INSPECTION RECORD INDICATED THAT THE LOT MET RELEASE CRITERIA. THE CARTRIDGE FORCE READINGS WERE CONFIRMED TO BE WITHIN SPECIFICATIONS; NO FAILURES RELATED TO THE LOSS OF PRIME COMPLAINT WERE NOTED. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE (OCCLUSION AND LOSS OF PRIME), CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOSS OF PRIME) ISSUE. REPORTEDLY, THE PUMP EMITTED MULTIPLE LOSS OF PRIME ALERTS. THE REPORTER STATED THAT THERE WAS NO SUDDEN CHANGE IN FORCE OR TEMPERATURE AND THE CARTRIDGE CAP WAS SECURED. REVIEW OF CARTRIDGE USAGE TECHNIQUES INDICATED THAT THE INSTRUCTION FOR USE WAS FOLLOWED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 251696 | ANIMAS INSULIN CARTRIDGE | INSULIN PUMP CARTRIDGE | LZG | ANIMAS CORPORATION | D200056 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 12 YR |