FDA Adverse Event Malfunction Summary report: N

9600

MDR report key: 3772062 · Received January 2, 2014

Report

Report Number
1720753-2014-00041
Event Type
Malfunction
Date Received
January 2, 2014
Date of Event
September 4, 2013
Report Date
January 2, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE SRAM BATTERY WAS EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SYSTEM FAILED TO BOOT UP. THERE ARE NO REPORTS OF PT INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2566 9600 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9600

Patients

Seq Age Sex Outcome Treatment
1