FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3771985 · Received December 30, 2013

Report

Report Number
8020893-2013-03288
Event Type
Malfunction
Date Received
December 30, 2013
Date of Event
January 1, 2013
Report Date
December 3, 2013
Manufacturer
COVIDIEN, FORMERLY NELLCOR
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE 840 VENTILATOR HAD BLANK BOTH DISPLAYS. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME THE MALFUNCTION OCCURRED. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE GRAPHICAL USER INTERFACE (GUI) PRINTED CIRCUIT BOARD (PCB) AND BACKLIGHT INVERTER PRINTED CIRCUIT BOARDS (PCB). THE UNIT PASSED EXTENDED SELF-TESTING.