FDA Adverse Event Injury Summary report: N

OT ULTRA2 METER

MDR report key: 3771972 · Received April 25, 2014

Report

Report Number
2939301-2014-09625
Event Type
Injury
Date Received
April 25, 2014
Date of Event
April 14, 2014
Report Date
April 16, 2014
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K053529
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 ¿ (08/04/2014). THE PATIENT¿S METER HAS BEEN RETURNED ON 7/21/2014 AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 7/22/2014 WITH THE FOLLOWING FINDINGS:THE METER PASSED ALL TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE REPRODUCED. THE RETAIN TEST STRIPS ALSO PASSED ALL TESTING. A DHR (DEVICE HISTORY RECORD) WAS COMPLETED ON 06/13/2014 FOR THIS PRODUCT AND NO DEVIATIONS, NON-CONFORMANCES, OR REWORKS WERE OBSERVED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Additional Manufacturer Narrative · 1

FOLLOW-UP # 2 ¿ (08/21/2014). THE PATIENT¿S TEST STRIPS HAVE BEEN RETURNED ON (B)(4) 2014 AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE TEST STRIPS PASSED ALL TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE CONFIRMED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Description of Event or Problem · 1

ON (B)(6) 2014, THE LAY USER/ PATIENT CONTACTED LIFESCAN (LFS) ALLEGING HIS ONETOUCH ULTRA2 METER READ INACCURATELY HIGH COMPARED TO ANOTHER DEVICE. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE ADVOCATE (CCA) DOCUMENTATION. THE ALLEGED ISSUE BEGAN IN THE EARLY MORNING OF (B)(6) 2014. THE PATIENT REPORTED BLOOD GLUCOSE RESULTS OF ¿198 MG/DL¿ WITH THE SUBJECT METER AND ¿171 MG/DL¿ ON ANOTHER METER, PERFORMED WITHIN 30 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS DO NOT EXCEED THE EXPECTED VALUE OF <= 30% AND/OR <= 30 MG/DL. THE PATIENT MANAGES HIS DIABETES WITH LANTUS INSULIN AND NOVOLOG INSULIN. THE PATIENT REPORTEDLY CONTINUED TO ADMINISTER HIS USUAL DOSE OF MEDICATIONS AT THAT TIME. ABOUT 4 HOURS LATER, THE PATIENT CLAIMED HE FELT SYMPTOMS OF SHAKEY AND SWEATING. THE PATIENT WAS GIVEN FOOD AND/ OR DRINK AS TREATMENT. AT THE TIME OF TROUBLESHOOTING, THE CCA NOTED THE SUBJECT METER WAS SET TO THE CORRECT UNIT OF MEASUREMENT AND AN APPROVED SAMPLE SITE WAS USED FOR TESTING. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT REPORTEDLY DEVELOPED SYMPTOMS SUGGESTIVE OF A SERIOUS INJURY AFTER THE ALLEGED METER ISSUE BEGAN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
251807 OT ULTRA2 METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3521053

Patients

Seq Age Sex Outcome Treatment
1 71 YR Life Threatening| R