FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 800 SYNCHRON® SYSTEM

MDR report key: 3771923 · Received April 25, 2014

Report

Report Number
2050012-2014-00214
Event Type
Malfunction
Date Received
April 25, 2014
Date of Event
April 5, 2014
Report Date
April 5, 2014
Manufacturer
BECKMAN COULTER
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A BECKMAN COULTER (BEC) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ASSESS THE INSTRUMENT. THE FSE FOUND A LOOSE REAGENT SYRINGE PLUNGER. THE FSE REPLACED THE BARREL ON THE LOOSE REAGENT SYRINGE. THERE WERE NO FURTHER REPORTS OF LEAKS OR OF OUT OF RANGE RESULTS DURING QUALITY CONTROL (QC.)

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT A UNICEL DXC 800 SYNCHRON SYSTEM PRODUCED BOTH HIGH AND LOW RESULTS FOR MULTIPLE CARTRIDGE REAGENTS DURING A QUALITY CONTROL (QC) RUN. THE CUSTOMER ALSO REPORTED LIQUID UNDER A REAGENT PROBE. THE LEAK WAS CONTAINED TO THE INSTRUMENT. THE OPERATOR WORE GLOVES AND A GOWN WHILE OPERATING THE INSTRUMENT. THERE WERE NO REPORTS OF DIRECT CONTACT WITH THE LEAK OR OF INJURY IN CONNECTION WITH THE LEAK. THE CUSTOMER REPORTS THAT TEN PATIENT RESULTS FOR MULTIPLE ANALYTES WERE AFFECTED. THESE RESULTS WERE RELEASED FROM THE LABORATORY AND WERE LATER AMENDED. THE CUSTOMER STATED THERE WAS NO CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT. PATIENT DATA WAS NOT PROVIDED BY THE CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
251207 UNICEL® DXC 800 SYNCHRON® SYSTEM ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE) FOR CLINICAL USE JJE BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1