FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3771828 · Received April 24, 2014

Report

Report Number
2531779-2014-11274
Event Type
Malfunction
Date Received
April 24, 2014
Report Date
April 10, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 04/10/2014 WITH THE FOLLOWING FINDINGS: ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. DURING TESTING, THE OK AND CONTRAST KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE AND REQUIRED MULTIPLE PRESSES TO ENGAGE; ALL OTHER KEYPAD BUTTONS RESPONDED PROPERLY. THE KEYPAD COVER WAS REMOVED AND CONTAMINATION WAS FOUND UNDER ALL BUTTON CONTACTS. ADDITIONALLY, THE DISPLAY SCREEN TEXT APPEARED DIM, FADED, AND DISCOLORED. THE BATTERY COMPARTMENT WAS ALSO OBSERVED TO BE CRACKED. UNRELATED TO THESE ISSUES, THE KEYPAD COVER WAS FOUND TO BE TORN OFF BELOW THE OK BUTTON, EXPOSING A BUTTON CONTACT, AND A HOLE WAS OBSERVED IN THE BOLUS BUTTON COVER. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A KEYPAD BUTTON RESPONSE ISSUE, CONTAMINATION UNDER KEYPAD BUTTONS, A DISPLAY FAILURE, AND A BATTERY COMPARTMENT CRACK. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 04/10/2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
249719 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 56 YR