PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3771764 · Received April 24, 2014

Report

Report Number
3004209178-2014-83936
Event Type
Malfunction
Date Received
April 24, 2014
Date of Event
February 27, 2014
Report Date
April 17, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP ALARMED DURING THE BASIC OCCLUSION TEST AS A RESULT OF A PROTRUDED/LOOSE DRIVE SUPPORT DISK. THE DEVICE WAS RECEIVED WITH CRACKED CASE AT THE DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE AND LIP, AND BROKEN BELT CLIP SLOT, SCRATCHED SCREEN, CRACKED BATTERY TUBE THREADS, AND BROKEN BELT CLIP SLOT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP KEPT ALARMING WHILE PRIMING AND INSULIN SQUIRTED OUT. ADVISED THE CALLER THAT THE INSULIN PUMP WOULD BE REPLACED. NO FURTHER INFORMATION WAS PROVIDED.