FDA Adverse Event
Malfunction
Summary report: N
PARADIGM INSULIN INFUSION PUMP
MDR report key: 3771739
·
Received April 24, 2014
Report
- Report Number
- 3004209178-2014-83897
- Event Type
- Malfunction
- Date Received
- April 24, 2014
- Date of Event
- November 19, 2013
- Report Date
- April 15, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- LZG
- PMA / PMN Number
- K031390
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP FAILED THE DISPLACEMENT TEST AND ALARMED MOTOR ERROR DURING REWIND AS A RESULT OF A MOTOR ENCODER SIGNAL BEING OUT OF PHASE. THE DEVICE HAD CRACKED RESERVOIR TUBE LIP.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP ALARMED MOTOR ERROR WHILE REWINDING AND THE ALARM HISTORY CONFIRMED THE ALARM. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 249656 | PARADIGM INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-712WWL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |